750wb field info:

Date 2H Drift(units/hr) BF(MHz) 2H_field 1H drift(hz/hr)
12/2013 -37.7 750.15 ? ?
07/2014 -40.2 750.10 ? ?
04/2015 -34.19 750.05 ? ?
10/2016 -34.3 749.95 ? ?
10/2017 ? 749.89 -200 ?
04/2018 ? 749.89 -5748 ?
08/2018 -35.5 749.89 -9897 -5.5
08/2018 -35.5 749.845 -560 -5.5
11/14/2018 -35.5 749.845 -3860 -
5/1/2019 -35.5 749.845 -9763 -
5/1/2019 -35.5 749.800 -200 -
2/11/2020 -35.5 749.750 762 -
12/09/2020 -35.5 749.700 662 -
10/24/2021 -35.5 749.650 148 -
11/10/2021 ssnmr 749.653631 onres adamantane 38.48ppm -

The 2H drift ~ -10000/9months (field units/time). This equates to roughly -38kHz in terms of proton drift. Need to update the base frequency every 9 months approximately through 'CF'. Also, be sure to update "expinstall" for default parameters for topshim to work etc.

solids mas controller: tty = 04
- use high power 15N filter (ishii labelled). 13C/1H are the same as liquids